Preliminary Results with the Shape Memory Nail: A Self-contained Distal Locking Mechanism for Diaphyseal Femur Fractures


Nando Ferreira1, Luan Nieuwoudt2

1Division of Orthopaedic Surgery, Department of Surgical Sciences, Faculty of Medicine and Health Sciences, Stellenbosch University, Cape Town, South Africa
2Department of Orthopaedic Surgery, University of Kwa-Zulu Natal, Pietermaritzburg, South Africa

Corresponding Author: Nando Ferreira, Division of Orthopaedic Surgery, Department of Surgical Sciences, Faculty of Medicine and Health Sciences, Stellenbosch University, Cape Town, South Africa, Phone: +27833245585, e-mail: drferreiran@gmail.com

How to cite this article Ferreira N, Nieuwoudt L. Preliminary Results with the Shape Memory Nail: A Self-contained Distal Locking Mechanism for Diaphyseal Femur Fractures. Strategies Trauma Limb Reconstr 2019;14(3):115–120.

Source of support: Nil

Conflict of interest: None

ABSTRACT

Distal interlocking of intramedullary nails can be challenging if not done regularly and can be associated with a prolonged operating time and excessive radiation exposure. Multiple techniques have been developed to overcome these problems but all still rely on conventional distal locking methods. Between December 2011 and March 2013, 18 patients with diaphyseal femur fractures were treated with the shape memory nail (Orthofix, Verona, Italy). These nails use self-contained nitinol memory metal ‘wings’ at the distal aspect of the nail to provide rotational and longitudinal stability. We observed fracture union in all 18 cases with no non-unions, rotational malalignments or peri-prosthetic infections. Median theatre time was 35 (18–71) minutes and median total radiation time was 50 (20–209) seconds. The shape memory nail (Orthofix, Verona, Italy) is an attractive alternative to conventional interlocking femoral nails. It provides sufficient stability to allow fracture union while decreasing theater time and limiting radiation exposure.

Keywords: Distal locking, Femur fracture, Intramedullary nail, Memory metal, Nitinol.

INTRODUCTION

Interlocked intramedullary nails are a gold standard in the operative management of adult diaphyseal femur fractures.1 These nails rely on interlocking screws to provide rotational and longitudinal stability during fracture healing. Inserting these distal interlocking screws can be technically demanding if not done regularly, and can be associated with prolonged operative time and increased radiation exposure.2

Traditional distal locking screw insertion relies on intraoperative X-ray imaging to guide placement of these screws. Although the design of intramedullary nails has improved significantly, most still rely on traditional interlocking screws for rotational and longitudinal stability.

Multiple techniques and devices have been developed to assist with distal locking including free-hand techniques, nail mounted guides, image intensifier mounted guides and computer assisted techniques.310 Keenan et al. described the McIndoe scissor technique to help locate both distal locking holes simultaneously.11 Owen et al. evaluated a K-wire and cannulated drill technique and found no benefit when compared to the free-hand technique.6 Rohilla et al. and Soni et al. described the nail-over-nail technique to simplify distal locking and decrease radiation exposure.7,12 Nail mounted jig techniques have been developed by multiple companies, including Orthofix (Verona, Italy), Stryker (Kalamazoo, Michigan), Synthes (Zuchwil, Switzerland) and the surgical implant generation network (SIGN) nail.1318 The inherent problem with these nail-mounted techniques is the slight deformation that femoral nails undergo during insertion into the intramedullary canal.19 Goodall et al. and Goulet et al. described image intensifier mounted lasers that allow aiming of the drill through the hole.8,20 The Trigen Sureshot (Smith and Nephew, Memphis, Tennessee) relies on electromagnetic computer-assisted navigation to insert distal locking screws.3,21 More recently, robot navigation systems have been developed to assist with distal locking.22,23 All of these techniques, however, still rely on conventional interlocking screws to provide rotational and longitudinal stability.

We report our preliminary results of diaphyseal femur fractures treated with the Orthofix Shape Memory Nail (Orthofix, Verona, Italy). This nail contains Nitinol memory metal ‘wings’ to provide rotational and longitudinal stability instead of conventional interlocking screws.

MATERIALS AND METHODS

Between December 2011 and March 2013 all patients who presented with isolated diaphyseal femur fractures and were treated with the Orthofix Shape Memory Nail (SMN) were eligible for inclusion. Patients were excluded if they did not complete at least 12 months of follow-up after the index procedure. Institutional ethics committee approval was obtained for this study.

All patients were operated on a fracture table and with the use of a fluoroscopic C-arm. An incision proximal to the greater trochanter allowed access to the tip of the greater trochanter. After proximal entry reaming, a guide wire was inserted and used to measure the appropriate nail length. The ideal nail depth was measured to be 20 mm from the intercondylar notch. Femoral canal preparation followed with reaming up to size 13 reamers.

After selection of the appropriate length nail, the nail was attached to the jig and the nitinol wings closed as per the described surgical technique. Nail insertion to the appropriate depth was followed by controlled deployment of the Nitinol ‘wings’ (Fig. 1). Two proximal interlocking screws and a nail end cap completed the surgical procedure. After wound closure, the patient was returned to the ward and received 24 hours of prophylactic antibiotics (1 g Cephazolin 8 hourly). Surgical time from skin incision to skin closure was recorded for all surgeries. The radiation exposure, in seconds, for each case was read from the fluoroscopy machine and recorded. Total radiation exposure during nail insertion, Nitinol wing deployment, and proximal screw placement was recorded separately.

Functional rehabilitation was encouraged under the guidance of a physiotherapist with a program focussing on early mobilisation and weight bearing followed by normalisation of gait pattern and functional use. No period of restricted weight bearing was used. Outpatient follow-up was scheduled at 6 weeks; 3 months, 6 months and 12 months after nail insertion. Nail removal was considered after 12 months and after radiographic evidence of fracture union. Nail removal was performed with patients in the lateral position. The proximal end of the nail was exposed followed by end cap and proximal locking screw removal. The nail removal jig was attached to facilitate closure of the Nitinol wings followed by removal of the nail. After nail removal, patients were allowed to continue functional use of the limb without any period of restricted activity.

RESULTS

Eighteen patients met the inclusion criteria. No patients were excluded. The cohort comprised 14 men and four women with a mean age of 29.7 years (range 18–56, SD 10.4) (Table 1). Mean follow-up was 13.4 months (range 12–18). In total, 11 fractures were classified as AO 32A, six OA 32B and one AO 32C.

Medical co-morbidities were identified in four patients (27%). These four patients were all HIV positive with cluster of differentiation 4 (CD4) counts ranging from 192 to 584 cells/mm3 (mean = 381.5 cells/mm3). All four patients were on highly active anti-retroviral (HAART) treatment. Four patients (27%) were active smokers.

Surgical time ranged from 18 to 71 minutes (median = 35, SD 18) and showed a significant variance between different surgeons. Additionally, a gradual reduction in operative time was observed with subsequent surgeries indicating the learning curve associated with the use of a new device (Fig. 2).

Fig. 1: AP and lateral radiographs showing deployment of Nitinol ‘wings’

Fig. 2: Time (minutes) per case for each surgeon

Table 1: Details of results
PatientAgeGenderAONail lengthSurgical time (min)Radiation time: nail insertion (seconds)Radiation time: wings opening (seconds)Radiation time: proximal screws (seconds)Radiation time: total (seconds)Callus
  139M32A39540164430Hypertrophic
  218M32A41025234945Hypertrophic
  356M32B35024144543Hypertrophic
  419M32A41035141654Hypertrophic
  530M32B39527351945Normotrophic
  632M32B39528171934Normotrophic
  720M32A410682437203Normotrophic
  830F32C38029213651Hypertrophic
  932F32A3953618810117Hypertrophic
1019M32A39530254858Hypertrophic
1135M32A3807114117135Normotrophic
1233M32A38062992114Normotrophic
1322F32B38030253432Normotrophic
1420M32A410716321378Normotrophic
1519M32A39541398350Normotrophic
1646F32A38051255541Hypertrophic
1732M32B39523101720Normotrophic
1833M32B39518121625Normotrophic
Median30   3519.53.56.550 
Mean29.7   39.4224765 
STDEV10.4   18133348 
Range18–56   18–719–631–113–1320–135 

Fig. 3: Radiation time (seconds) per case for each surgeon

Fig. 4: AP and lateral radiographs showing union of a midshaft femur fracture 12 months after nail insertion

Fig. 5: AP and lateral radiographs showing hypertrophic callus formation 6 weeks after nail insertion

Total fluoroscopic screening time, similarly, showed a significant variance between surgeons. Screening time ranged from 20 seconds to 209 seconds (median = 50, SD 63). Here too, screening time tended to decrease with subsequent cases, though not as marked as seen with surgical time (Fig. 3). The mean screening time for distal locking through opening of the Nitinol wings were 4 seconds and ranged from 1 to 11 seconds.

All 18 fractures united without any additional surgeries or adjuvant therapies. Sixteen of the 18 fractures (88.9%) were clinically and radiologically united at the 3-month follow-up. The two remaining patient’s fractures were confirmed to be united at their 6-month follow-up. Ten patients (55.6%) united with normotrophic callus (Fig. 4) while the remaining eight patients (44.4%) had hypertrophic callus (Fig. 5) evident on follow-up radiographs. No superficial or deep infection developed. No malunions occurred and no patient reported distal femur pain.

Six patients agreed to elective nail removal (Fig. 6). Four patients refused removal and the remaining eight patients never returned after their 12-month follow-up. The removals were performed according to the described surgical technique and were accomplished without incident.

DISCUSSION

The aim of this study was to present our initial experience of diaphyseal femur fractures treated with the Orthofix Shape Memory Nail (Orthofix, Verona, Italy). The unique Nitinol memory metal distal locking mechanism provided an alternative to conventional distal interlocking screws.

Fig. 6: AP and lateral radiographs of femur after removal of nail

Nitinol stands for nickel (Ni), titanium (Ti) and Naval Ordnance Laboratory (NOL) where it was discovered in the early 1960s and used in medical devices since the 1970s.2427 It is the generic trade name for the nickel and titanium alloy where the two metals are present in near equal anatomic percentages. Nitinol exhibits two unique properties namely shape memory effect and superelasticity. The shape memory effect refers to the alloys ability to be deformed at certain temperatures ranges and then return to its original shape when heated. It is this property that is frequently exploited for medical purposes as is the case in the shape memory nail (Orthofix, Verona, Italy).

In addition to the Nitinol distal locking mechanism in the SMN by Orthofix, several other companies have developed solutions to assist surgeons to achieve length and rotational stability following long-bone intramedullary nailing. Expandable self-locking nails have shown initial promise but reports of implant failure and limb shortening have curtailed their use.2833 The electromagnetic computer-assisted Trigen Sureshot (Smith and Nephew, Memphis, Tennessee) system retains the traditional interlocking screws but uses real-time three-dimensional ‘navigation’ to assist with distal locking without the need for fluoroscopic radiation.3,21

In addition to finding solutions to simplify distal locking, strategies to reduce radiation exposure during distal locking is also a priority. Krettek et al. compared the use of a targeting device for distal locking with the free hand technique.5 They reported that although the targeting device was slower than the free-hand technique for distal locking (6.6 ± 2.4 minutes vs 4.8 ± 1.5 minutes), total radiation time was significantly less with use of the targeting device (23 ± 17 seconds vs 69 ± 34 seconds). The authors reported total radiation time for distal locking only using the free-hand technique to be 37 ± 15.5 seconds.5 Hoffmann et al. reported a median time benefit of 244 seconds without using ionising radiation when comparing distal locking with conventional fluoroscopic guidance and Sureshot (Smith and Nephew, Memphis, Tennessee) electromagnetic-guided methods in 50 cadavers.3 Similarly, Horn et al. reported a significant reduction in radiation exposure (230.54 μGy vs 690.27 μGy) when comparing Sureshot with free-hand distal locking.34 Theoretically, apart from screening for confirmation, no fluoroscopic imaging is required during the deployment of the Nitinol distal locking ‘wings’. This was illustrated in our findings with a mean time for fluoroscopic screening during deployment being 4 seconds, but being as low as 1 second in multiple cases.

The low number of cases enrolled during this pilot study is a clear limitation. The fact that only six nails were removed during the study period is also a concern; although all of these removals were uneventful, the ability to close the Nitinol wings for extraction need more rigorous testing.

CONCLUSION

The Orthofix Shape Memory Nail is an attractive alternative to conventional interlocking femoral nails. The nitinol memory metal distal locking ‘wings’ provide sufficient rotational and longitudinal stability to allow fracture union while limiting radiation exposure.

REFERENCES

1. Webb LX, Gristina AG, Fowler HL. Unstable femoral shaft fractures: a comparison of interlocking nailing versus traction and casting methods. J Orthop Trauma 1988;2(1):10–12. DOI: 10.1097/00005131-198802000-00003.

2. Sugarman ID, Adam I, Bunker TD. Radiation dosage during AO locking femoral nailing. Injury 1988;19(5):336–338. DOI: 10.1016/0020-1383(88)90107-6.

3. Hoffmann M, Schroder M, Lehmann W, et al. Next generation distal locking for intramedullary nails using an electromagnetic X-ray-radiation-free real-time navigation system. J Trauma Acute Care Surg 2012;73(1):243–248. DOI: 10.1097/TA.0b013e31824b0088.

4. Krettek C, Könemann B, Miclau T, et al. A new mechanical aiming device for the placement of distal interlocking screws in femoral nails. Arch Orthop Trauma Surg 1998;117(3):147–152. DOI: 10.1007/s004020050215.

5. Krettek C, Konemann B, Miclau T, et al. A mechanical distal aiming device for distal locking in femoral nails. Clin Orthop Relat Res 1999(364):267–275. DOI: 10.1097/00003086-199907000-00033.

6. Owen TD, Coorsh J. Insertion of the distal locking screws in femoral nailing: a simplified technique. Injury 1993;24(2):101–103. DOI: 10.1016/0020-1383(93)90198-f.

7. Rohilla R, Singh R, Magu N, et al. Nail over nail technique for distal locking of femoral intramedullary nails. Int Orthop 2009;33(4):1107–1112. DOI: 10.1007/s00264-008-0579-y.

8. Goodall JD. An image intensifier laser guidance system for the distal locking of an intramedullary nail. Injury 1991;22(4):339. DOI: 10.1016/0020-1383(91)90026-b.

9. Arlettaz Y, Dominguez A, Farron A, et al. Distal locking of femoral nails: evaluation of a new radiation-independent targeting system. J Orthop Trauma 2012;26(11):633–637. DOI: 10.1097/BOT.0b013e318242d8b7.

10. Bonshahi AY, Cowey A, Vhadra R. A simple distal interlocking aid for intramedullary nails. J Postgrad Med 2003;49(1):99–100. DOI: 10.4103/0022-3859.912.

11. Keenan J, Stitson D, Charnley GJ. McIndoe’s scissors - a simple technique to aid distal locking of an intramedullary nail. Ann R Coll Surg Engl 2000;82(2):143–144.

12. Soni R, Mehta S, Singh J, et al. Radiation-free insertion of distal interlocking screw in tibial and femur nailing: a simple technique. J Surg Tech Case Report 2012;4(1):15–18. DOI: 10.4103/2006-8808.100346.

13. Anastopoulos G, Ntagiopoulos PG, Chissas D, et al. Distal locking of tibial nails: a new device to reduce radiation exposure. Clin Orthop Relat Res 2008;466(1):216–220. DOI: 10.1007/s11999-007-0036-z.

14. Krettek C, Konemann B, Farouk O, et al. Experimental study of distal interlocking of a solid tibial nail: radiation-independent distal aiming device (DAD) versus freehand technique (FHT). J Orthop Trauma 1998;12(6):373–378. DOI: 10.1097/00005131-199808000-00001.

15. Babis GC, Benetos IS, Karachalios T, et al. Eight years’ clinical experience with the orthofix tibial nailing system in the treatment of tibial shaft fractures. Injury 2007;38(2):227–234. DOI: 10.1016/j.injury.2006.07.030.

16. Pennig D, Oppenheim W, Faccioli G, et al. Intramedullary locked nailing of femur and tibia: insertion of distal locking screws without image intensifier. Injury 1997;28(4):323–326. DOI: 10.1016/s0020-1383(97)00004-1.

17. Chakraborty MK, Thapa P, Sathian B. Surgical implant generation network (SIGN) solid intramedullary interlocking nail in the lower extremity: an observational study from western Nepal. J Clin Diag Res 2011;5(8):1614–1617.

18. Pardiwala D, Prabhu V, Dudhniwala G, et al. The AO distal locking aiming device: an evaluation of efficacy and learning curve. Injury 2001;32(9):713–718. DOI: 10.1016/s0020-1383(01)00100-0.

19. Krettek C, Mannss J, Miclau T, et al. Deformation of femoral nails with intramedullary insertion. J Orthop Res 1998;16(5):572–575. DOI: 10.1002/jor.1100160508.

20. Goulet JA, Londy F, Saltzman CL, et al. Interlocking intramedullary nails. n improved method of screw placement combining image intensification and laser light. Clin Orthop Relat Res 1992(281):199–203.

21. Stathopoulos I, Karampinas P, Evangelopoulos DS, et al. Radiation-free distal locking of intramedullary nails: evaluation of a new electromagnetic computer-assisted guidance system. Injury 2013;44(6):872–875. DOI: 10.1016/j.injury.2012.08.051.

22. Shoham M, Burman M, Zehavi E, et al. Bone-mounted miniature robot for surgical procedures: concept and clinical applications. IEEE Trans Rob Autom 2003;19(5):893–901. DOI: 10.1109/TRA.2003.817075.

23. Yaniv Z, Joskowicz L. Precise robot-assisted guide positioning for distal locking of intramedullary nails. IEEE Trans Med Imaging 2005;24(5):624–635. DOI: 10.1109/TMI.2005.844922.

24. Pelton AR, Stockel D. A guide to shape memory and superelasticity in nitinol medical devices. Minim Invasive Ther Allied Technol 2004;13(4):218–221. DOI: 10.1080/13645700410017236.

25. Stoeckel D, Pelton A, Duerig T. Self-expanding nitinol stents: material and design considerations. Eur Radiol 2004;14(2):292–301. DOI: 10.1007/s00330-003-2022-5.

26. Duerig T, Pelton A, Stockel D. An overview of nitinol medical applications. Mater Sci Eng 1999(A273–275):149–160. DOI: 10.1016/S0921-5093(99)00294-4.

27. Cutright DE, Bhaskar SN, Perez B, et al. Tissue reaction to nitinol wire alloy. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 1973;35(4):578–584. DOI: 10.1016/0030-4220(73)90017-0.

28. Rose DM, Smith TO, Nielsen D, et al. Expandable intramedullary nails in lower limb trauma: a systematic review of clinical and radiological outcomes. Strat Traum Limb Recon 2013;8(1):1–12. DOI: 10.1007/s11751-013-0156-9.

29. Kapoor SK, Kataria H, Boruah T, et al. Expandable self-locking nail in the management of closed diaphyseal fractures of femur and tibia. Indian J Orthop 2009;43(3):264–270. DOI: 10.4103/0019-5413.53457.

30. Sipahioglu S, Zehir S, Sarikaya B, et al. Comparision of the expandable nail with locked nail in the treatment of closed diaphyseal fractures of femur. Niger J Clin Pract 2017;20(7):792–798. DOI: 10.4103/1119-3077.212452.

31. Smith WR, Ziran B, Agudelo JF, et al. Expandable intramedullary nailing for tibial and femoral fractures: a preliminary analysis of perioperative complications. J Orthop Trauma 2006;20(5):310–314. DOI: 10.1097/00005131-200605000-00002 ;discussion 5-6.

32. Maugans CJ, Damron TA. Femoral fixion expandable nail implant failure and difficult extraction: a case report. Tech Orthop 2015;30(1):e1–e4. DOI: 10.1097/BTO.0000000000000087.

33. Lepore L, Lepore S, Maffulli N. Intramedullary nailing of the femur with an inflatable self-locking nail: comparison with locked nailing. J Orthop Sci 2003;8(6):796–801. DOI: 10.1007/s00776-003-0709-8.

34. Horn A, Maqungo S, Roche S, et al. Distal locking in femoral intramedullary nails: free-hand technique versus Sureshot® assisted. Orthop Proc 2013;95-B (Supp. 14):20.